The instrument uses sixteen cadmium telluride (CdTe) detectors with iridium electrodes that act as Schottky barriers. Although the instrument operated properly, the instrument power supply failed at 9:45 JST. The instrument was successfully recovered. The cause of the fault was determined to be inadequate battery supply leading to excessive heating due to the high current draw, triggering the thermal fuse.
Balloon launched on: 8/29/2001 at 6:30 jst
Launch site: Sanriku Balloon Center, Iwate, Japan
Balloon launched by: Institute of Space and Astronautical Science (ISAS)
Balloon manufacturer/size/composition: Zero Pressure Balloon model B80 80.000 m3
Flight identification number: B80-6
End of flight (L for landing time, W for last contact, otherwise termination time): 8/29/2001
Balloon flight duration (F: time at float only, otherwise total flight time in d:days / h:hours or m:minutes - ): 5 h 7 m